Case report: The membrane oxygenator failure during cardiopulmonary bypass for cardiac surgery - a rare incident

  • Nguyễn Đăng Thứ Bệnh viện Quân y 103
  • Trần Đắc Tiệp Bệnh viện Quân y 103
  • Khuất Đình Huyến Bệnh viện Quân y 103
  • Nguyễn Ngọc Trung Bệnh viện Quân y 103
  • Vũ Đức Thắng Bệnh viện Quân y 103
  • Lê Bá Hạnh Bệnh viện Quân y 103
  • Nguyễn Thế Kiên Bệnh viện Quân y 103

Main Article Content

Keywords

Cardiopulmonary bypass, oxygenator failure

Abstract

Cardiopulmonary bypass (CPB) is more and more developed and is an essential component of modern cardiac surgery. The advanced technology of the CPB machine and oxygenator help the CPB procedure to be reliable and accurate. However, due to the complexity of the CPB system, equipment failure can potentially occur. We report a case in which a membrane oxygenator failure was discovered during initiating CPB. Quickly assessing and smooth replacement of the oxygenator is to help treat a patient in the safest possible setting. Simulation practices are important in improving efficiency in emergency situations. We propose that all perfusionists should be aware of these incidents and an algorithm should be put in place and adhered to with regular simulation drills to improve efficiency if such a situation were to occur.

Article Details

References

1. Iwahashi H, Yuri K, Nosé Y (2004) Development of the oxygenator: Past, present, and future. J Artif Organs 7(3): 111-120.
2. Palanzo DA (2005) Perfusion safety: Defining the problem. Perfusion 20(4): 195-203.
3. Soo A, Booth K, Parissis H (2012) Successful management of membrane oxygenator failure during cardiopulmonary bypass the importance of safety algorithm and simulation drills. J Extra Corpor Technol 44(2): 78-80.
4. Fernandes A, Laliberte E, Toledano K et al (2015) A novel method to detect an oxygenator defect prior to cardiopulmonary bypass initiation. J Extra Corpor Technol 47(3): 180-182.
5. Groom RC, Forest RJ, Cormack JE et al (2002) Parallel replacement of the oxygenator that is not transferring oxygen: The PRONTO procedure. Perfusion 17(6): 447-450.
6. Darling E, Searles B (2010) Oxygenator change-out times: The value of a written protocol and simulation exercises. Perfusion 25(3): 141-143; discussion 144-145.